BARD PERIPHERAL VASCULAR, INC. ECLIPSE FILTER SYSTEM - JUGULAR; VENA CAVA FILTER Back to Search Results
Catalog Number EC500J
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Thrombosis (2100)
Event Date 03/27/2012
Event Type  Injury  
Manufacturer Narrative
Manufacturing review: a lot history review was performed.This is the only complaint to date for this lot number.Therefore, a device history record review is not required.Investigation summary: the device was not returned for evaluation.Medical records were provided and reviewed.There were no device deficiencies identified within the medical records.Therefore, the investigation is inconclusive as no objective evidence has been provided to confirm any alleged deficiency with the filter.Additionally, it can be confirmed that the patient experienced thrombus above the filter post deployment.However, the relationship to the filter is unknown.Based on the available information, the definitive root cause is unknown.Labeling review: a review of product labeling documents (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, and unit label) showed that the product labeling is adequate.(expiry date: 01/2014).
 
Event Description
It was reported through the litigation process that a vena cava filter was placed in a patient after being diagnosed with deep vein thrombosis and pulmonary embolism.Approximately seven months post filter deployment, the patient reportedly experienced thrombosis.The device has not been removed and there were no reported attempts made to retrieve the filter.The current status of the patient is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECLIPSE FILTER SYSTEM - JUGULAR
Type of Device
VENA CAVA FILTER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury NY 12804
Manufacturer Contact
judith ludwig
1415 w. 3rd street
tempe, AZ 85281
4803032689
MDR Report Key11106629
MDR Text Key224766064
Report Number2020394-2020-06524
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K093659
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 01/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/04/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberEC500J
Device Lot NumberGFUL1930
Was Device Available for Evaluation? No
Date Manufacturer Received12/08/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
METHADONE, LYRICA, AND NUCYNTA; MOTRIN, COUMADIN, VALIUM
Patient Outcome(s) Other;
Patient Age36 YR
-
-