| Model Number |
1012-04-060 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
Fall (1848); Foreign Body Reaction (1868); Pain (1994); Joint Laxity (4526)
|
| Date of Event |
12/14/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the ceramic head was fractured.Patient fell.No surgical delay.The acetabular cup was determined to be well-fixed and was not revised.Tri-lock stem was revised due to a fractured trunnion.Doi: (b)(6) 2013, dor: (b)(6) 2020, left hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary:no device associated with this report was received for examination.Examination of a provided x-ray image does not depict a fracture of the femoral stem as alleged.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Search Alerts/Recalls
|