MAUDE Adverse Event Report: LIFESCAN EUROPE GMBH OT VERIO 2 METER; GLUCOSE MONITORING SYS/KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIFESCAN EUROPE GMBH OT VERIO 2 METER; GLUCOSE MONITORING SYS/KIT Back to Search Results
Lot Number 4647038
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problems Fatigue (1849); Dizziness (2194); Dysphasia (2195); Pallor (2468); Shaking/Tremors (2515); Cognitive Changes (2551)
Event Date 12/14/2020
Event Type  Injury  
Event Description
On (b)(6) 2020, the lay-user/patient contacted lifescan (lfs) (b)(4), alleging that his onetouch verio2 meter read inaccurately high compared to his feelings and/or normal readings.The complaint was classified based on the customer care agent (cca) documentation.The patient informed the cca that the alleged meter inaccuracy began around (b)(6) 2020.The patient reported obtaining alleged inaccurate high readings of ¿22.4, 18.3, 14.0, 14.5, 17.4, 17.0, 22.3, 18.4, 27.4, 16.0, 16.3, 20.5, 16.3, 16.9, 16.8, 22.5, 15.1, 15.4 and 19.8 mmol/l¿ with the subject meter on separate occasions.The patient manages his diabetes with a combination of humalog and lantus insulin on a self-adjusting dose based on food intake.It was not reported if the patient made any changes to his usual diabetes management regimen in response to the elevated results.The patient reported experiencing ¿hypoglycemia¿ with symptoms of ¿light-headedness, shakiness, cannot concentrate, cannot speak, very weak and very pale¿ almost every night since the alleged issue began; he was unable to recall exact dates/times.The patient reported treating himself with an energy drink in response to the symptoms.On (b)(6) 2020, the patient reported receiving advice from his diabetes nursing team to reduce his dose of insulin from 56 to 44 units.The patient denied using any other device to test his blood glucose.At the time of troubleshooting, the cca confirmed the unit of measure was set correctly on the subject meter.The cca noted that the correct testing process was being followed.The cca confirmed the test strip vial was intact and the test strips had been stored correct and were not expired or opened past their discard date.The cca noted the patient did not have control solution available to perform a quality control test.Replacement product was sent to the patient.This complaint is being reported because the patient reportedly developed signs/symptoms suggestive of a serious injury adverse event after the alleged inaccuracy issue began.The subject meter could not be ruled out as a cause or contributor to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OT VERIO 2 METER
Type of Device
GLUCOSE MONITORING SYS/KIT
Manufacturer (Section D)
LIFESCAN EUROPE GMBH
gubelstrasse 34
zug 6300
SZ  6300
MDR Report Key11107991
MDR Text Key228179343
Report Number2939301-2021-02799
Device Sequence Number1
Product Code NBW
Combination Product (y/n)Y
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/04/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Lot Number4647038
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date12/22/2020
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
-
-