Model Number 406849 |
Device Problems
Fluid/Blood Leak (1250); Fracture (1260)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/07/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The results/method and conclusion codes along with investigation results will be provided in the final report.
|
|
Event Description
|
During an ablation procedure for atrial fibrillation, there was a crack on the tip of the outer sheath and blood leaked out of the introducer.The brk needle was inserted into the hemostasis valve prior to the reported leak.The device was replaced and the procedure was completed with no adverse consequences to the patient.
|
|
Manufacturer Narrative
|
One 8.5f fast-cath introducer sheath was received for evaluation.The dilator was also received.A leak was noted at the hemostasis valve during functional testing.The cap was removed from the hemostasis hub and the hemostasis seal was microscopically inspected.Tearing, resulting in a hole, was noted in slit of the seal.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the torn seal and subsequent leak is consistent with damage during use.The ifu states damage to the valve assembly may occur if inner catheter is withdrawn rapidly.
|
|
Manufacturer Narrative
|
One 8.5f fast-cath introducer sheath was received for evaluation.The dilator was also received.A leak was noted at the hemostasis valve during functional testing.The cap was removed from the hemostasis hub and the hemostasis seal was microscopically inspected.Tearing, resulting in a hole, was noted in slit of the seal.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the torn seal and subsequent leak is consistent with damage during use.
|
|
Search Alerts/Recalls
|