It was reported that when tunneling from superior neck to pocket subclavicular.The plastic piece inside the tunneller that you put the electrodes into broke in half by the leads as the doctor was pulling the tunneller out, thus leaving the leads half way under the skin in the neck.A small incision was made over the leads to retrieve them and the broken plastic, then they used another tunneller and attachment to tunnel the rest of the way.The healthcare provider (hcp) did have to pull rather hard because of anatomy.Once the leads were all connected to battery everything checked out good and the hcp was very happy.
|
|
Continuation of d10: product id 3555-60, lot# unknown, explanted: product type accessory product id 3555-60, lot# unknown, explanted: product type accessory, product id neu_carrier/2x4, lot# unknown, explanted: product type accessory, product id neu_carrier/2x4, lot# unknown, explanted: product type accessory.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|