MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL OG TDL CMPLX FRAME COIL 7X25; NEUROVASCULAR EMBOLIZATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL OG TDL CMPLX FRAME COIL 7X25; NEUROVASCULAR EMBOLIZATION DEVICE Back to Search Results
Model Number 640CR0725
Medical Device Problem Code Positioning Problem (3009)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/30/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Information regarding patient weight, height, medical history, race, and ethnicity was not reported.Procode: krd/hcg initial reporter phone: (b)(6).The device is available to be returned for evaluation and testing.However, it has not been received to date as indicated as ¿other¿ in this section as the reason for non-evaluation.If the device returns, a device investigation will be performed.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.The company is seeking this information through the event investigation.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.This is one of two products involved with the complaint and the associated manufacturer report numbers are 3008114965-2021-00001 and 3008114965-2021-00001.
 
Event or Problem Description
As reported by the field, during a coil embolization, it was reported that the coil of a 7x25 og tdl cmplx frame coil (640cr0725, 30396175) cannot be formed after push-out.It was reported that the coil cannot be formed after push-out.The physician changed the device to a 4x8 og cmplx xtrasoft coil (640cx0408, 30345369) and the same problem happened.The physician changed to another new one to complete the procedure.There was not patient injury report.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.Complaint conclusion: as reported by the field, during a coil embolization, it was reported that the coil of a 7x25 og tdl cmplx frame coil (640cr0725, 30396175) cannot be formed after push-out.The physician changed the device to a 4x8 og cmplx xtrasoft coil (640cx0408, 30345369) and the same problem happened.The physician changed to another new one to complete the procedure.There was not patient injury report.No additional information is available.A non-sterile unit og tdl cmplx frame coil 7x25 was received inside of a pouch.The device was inspected, and it was found in good normal conditions, no damages or anomalies were observed during the analysis.The device was inspected under a microscope and it was found that the embolic coil is separated from gripper inside the introducer, although the embolic coil was found separated from gripper no damages or anomalies were observed on it.A manufacturing record evaluation was performed for the finished device 30396175 number, and no non-conformances related to the reported complaint condition were identified.The complaint reported by the customer ¿coil - positioning difficulty-poor conformability¿ could not be duplicated because the complaint reported by the customer is specific to the patient and procedure at the time of occurrence and cannot be replicated in the lab.Initially, during the microscopic analysis, the embolic coil was inspected, and it was found in good normal conditions.However, further, along with the analysis, the embolic coil was found separated from gripper.This condition is not related with the customer¿s complaint regarding poor conformability.This condition could be originated due excessive force and/or handling applied to the device during the procedure time, therefore the event reported was not confirmed.Neither the analysis nor the mre suggests that the failure reported by the customer could not be related to the manufacturing process.The instructions for use (ifu) instructs on proper positioning of the microcoil system.The ifu also warns: ¿if repositioning of the microcoil is necessary, carefully observe the motion of the microcoil in respect to the dpu wire while retracting the microcoil under fluoroscopy.If the microcoil movement is not one-to-one with the dpu wire, or if repositioning is difficult, the microcoil may have become stretched and could possibly break.Gently remove and discard the microcoil system¿.Assignment of root cause for the event remains speculative and inconclusive, based on the limited information provided and the evidence presented by the returned device; however, it is possible that clinical and procedural factors, including device manipulation and device interaction, may have contributed to the reported failure and damages on the returned system.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OG TDL CMPLX FRAME COIL 7X25
Common Device Name
NEUROVASCULAR EMBOLIZATION DEVICE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle neuchatel CH-24 00
SZ  CH-2400
MDR Report Key11109393
Report Number3008114965-2021-00001
Device Sequence Number17124137
Product Code KRD
UDI-Device Identifier10886704030447
UDI-Public10886704030447
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K093973
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2022
Device Model Number640CR0725
Device Catalogue Number640CR0725
Device Lot Number30396175
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/13/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/11/2020
Supplement Date Received by Manufacturer01/13/2021
Initial Report FDA Received Date01/04/2021
Supplement Report FDA Received Date02/02/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
OG CMPLX XTRASOFT COIL 4X8
-
-