| Model Number |
980U1ENDIUU-LA |
| Medical Device Problem Code |
Smoking (1585)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/14/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic has not received the suspect device/component from the customer for evaluation nor has the device been evaluated by the service engineer.The distributor reported that the ventilator was evaluated on 2020-dec-14 and the user interface (ui) printed circuit board (pcb) was replaced.After replacement of the ui pcb, the internal battery failed which was replaced.The ventilator passed all testing and was in working condition after repair.If the replaced parts are returned for failure investigation, a supplemental medwatch report with the findings will be submitted once the investigation is complete.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that, when this 980 ventilator was powered on, the graphical user interface (gui) ¿appeared to have white smoke¿ and there was a ¿burnt component smell¿.The ventilator was not in use on a patient at the time of the reported event.
|
| |
|