Model Number 980U1ENDIUU-LA
Medical Device Problem Code Smoking (1585)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/14/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic has not received the suspect device/component from the customer for evaluation nor has the device been evaluated by the service engineer.The distributor reported that the ventilator was evaluated on 2020-dec-14 and the user interface (ui) printed circuit board (pcb) was replaced.After replacement of the ui pcb, the internal battery failed which was replaced.The ventilator passed all testing and was in working condition after repair.If the replaced parts are returned for failure investigation, a supplemental medwatch report with the findings will be submitted once the investigation is complete.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that, when this 980 ventilator was powered on, the graphical user interface (gui) ¿appeared to have white smoke¿ and there was a ¿burnt component smell¿.The ventilator was not in use on a patient at the time of the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
980 VENTILATOR
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
NELLCOR PURITAN BENNETT IRELAND
micheal collins rd mervue
galway
EI 
MDR Report Key11110864
Report Number8020893-2021-00003
Device Sequence Number3922940
Product Code CBK
UDI-Device Identifier10884521201392
UDI-Public10884521201392
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K131252
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number980U1ENDIUU-LA
Device Catalogue Number980U1ENDIUU-LA
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/14/2020
Initial Report FDA Received Date01/04/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-