MAUDE Adverse Event Report: MPRI SPECIFY SURESCAN; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI SPECIFY SURESCAN; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF Back to Search Results
Model Number 977C165
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Positioning Problem (3009)
Health Effect - Clinical Codes Unspecified Nervous System Problem (4426); Insufficient Information (4580)
Date of Event 12/30/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from multiple sources (manufacturer representative, healthcare provider) regarding a patient who was implanted with an implantable neurostimulator (ins).It was reported that the patient had a loss of motor testing signals for the right lower extremity during paddle placement.It was unknown what led to the issue.The paddle was removed and perfusion was increased intraoperatively.It was unknown if the issue was resolved.
 
Additional Manufacturer Narrative
H3: the returned lead passed all testing in the laboratory and no anomalies were identified.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Additional information was received from the manufacturing representative (rep) and it was reported that motor signals did not recover in surgery.However the patient did appear to have ability to move the right lower extremity.Paddle has been collected and will be returned.
 
Event or Problem Description
The lead was received for analysis on (b)(6) 2021.It was noted that there was positioning difficulty with the paddle lead and neurologic pathway changes after placement.There was patient injury; however, the patient recovered without sequela.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECIFY SURESCAN
Common Device Name
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key11111601
Report Number2649622-2021-00092
Device Sequence Number3608983
Product Code LGW
UDI-Device Identifier00643169464599
UDI-Public00643169464599
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
P840001
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 02/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/02/2024
Device Model Number977C165
Device Catalogue Number977C165
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/30/2020
Supplement Date Received by Manufacturer01/04/2021
01/22/2021
02/17/2021
Initial Report FDA Received Date01/04/2021
Supplement Report FDA Received Date01/05/2021
02/05/2021
02/18/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age60 YR
Patient Weight95
-
-