SMITH & NEPHEW, INC. REF XLPE 36 20 DEG 62-64 H; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
|
Back to Search Results |
|
| Model Number |
71333347 |
| Medical Device Problem Code |
Positioning Problem (3009)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/13/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the reflection xlpe liner did not lock in the shell.The procedure was completed using a s+n backup device.A surgical delay of 30 min or less was reported.No harm to the patient was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: the device, used in treatment, was returned for evaluation.A visual inspection was completed but could not confirm the state failure.There are gouges and minor damage to the liner.The liner exhibits deep scratches/deformation on the id and id rim.A dimensional analysis could not be completed due to damage to the liner from attempting to insert it into the shell.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.Some potential probable causes for this event could include a fit/ sizing issue or poor insertion technique.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|