U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. PICO 7 10CMX40CM; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. PICO 7 10CMX40CM; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP Back to Search Results
Model Number 66022004
Medical Device Problem Code Pumping Stopped (1503)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/11/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that during treatment the pico pump failed, with all 4 lights remaining on.A backup was available to continue treatment with a minimum delay of less than 30 minutes.
 
Additional Manufacturer Narrative
H3: h6: we have now completed our investigation into the reported complaint.A review of the batch manufacturing records could not be performed as no lot number was provided, however, there are no indications to suggest that the device did not meet specifications upon release into distribution.The complaint history file contains further instances/ related events of the reported event.As no product could be returned, a thorough evaluation of the device could not be carried out.An image provided confirmed a relationship between the event and the device.The device was intended for use in treatment.It was reported that all indicator lights were solidly illuminated.This means that the device entered a non-recoverable error state.There are various reasons that the device may enter an error state, such as out of range negative pressure, out of range power supply and an error retrieving data.If the pump does enter an error state, it must be replaced.This is a patient safety feature.We have not been able to identify a definitive root cause.No further actions by smith and nephew are deemed necessary at this stage.However, we will continue to monitor for any adverse trends relating to this product range.Smith and nephew acknowledge customer concern and are continually investigating ways to develop and improve our products.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PICO 7 10CMX40CM
Common Device Name
NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull HU3 2 BN
UK  HU3 2BN
MDR Report Key11112975
Report Number8043484-2021-00014
Device Sequence Number3698556
Product Code OMP
UDI-Device Identifier05000223492199
UDI-Public05000223492199
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K180698
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number66022004
Device Catalogue Number66022004
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/11/2020
Supplement Date Received by Manufacturer01/29/2021
Initial Report FDA Received Date01/04/2021
Supplement Report FDA Received Date02/03/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-