| Model Number |
95-0016 |
| Medical Device Problem Code |
Loss of or Failure to Bond (1068)
|
| Health Effect - Clinical Codes |
Pain (1994); Osteolysis (2377); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
12/17/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Patient underwent left primary tkr in 2006.At this time a primary pfc rpf femoral component size 5 left and a cementless duofix mbt tray was utilised (cat numbers and lot numbers could not be established).Patient has done well for a number of years (undertaking manual work) and started to feel pain in the last 2 years.X-rays show a loose femoral component along with wear of the tibial insert on the medial side.On opening the knee the femoral component was obviously loose (this component had debonded at the implant cement interface and showed signed of polishing where there had been movement between the cement and the implant).Granulomas soft tissue was excised from the extensor mechanism, and there were eteolytic lesions in the distal femoral bone.The duofix tray was not overtly loose but did come away from the tibial bone without too much effort.The patella component was left in situ.An attune revision component was implanted along with both femoral and tibial sleeves.Good stability and fixation were achieved.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot: dhr reviewed no deviations or non-conformances were noted.
|
| |
|
Search Alerts/Recalls
|