MAUDE Adverse Event Report: GLOBUS MEDICAL, INC. RESONATE; RESONATE ANTERIOR CERVICAL PLATE, 3-LEVEL, 42MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GLOBUS MEDICAL, INC. RESONATE; RESONATE ANTERIOR CERVICAL PLATE, 3-LEVEL, 42MM Back to Search Results
Model Number 1194.3042
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not available for evaluation as it remains in the patient.Imaging provided shows an inferior screw has backed out anteriorly past the screw blocking mechanism.The patient is reported to be asymptomatic and there are no plans for a revision surgery at this time.No determinations can be made as to the cause of the reported issue.
 
Event or Problem Description
It was reported that a screw backed out of the plate post-operatively.There are no plans for a revision surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESONATE
Common Device Name
RESONATE ANTERIOR CERVICAL PLATE, 3-LEVEL, 42MM
Manufacturer (Section D)
GLOBUS MEDICAL, INC.
2560 general armistead avenue
audubon 19403
MDR Report Key11113609
Report Number3004142400-2021-00006
Device Sequence Number3472116
Product Code KWQ
UDI-Device Identifier00193982155353
UDI-Public00193982155353
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K192314
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 12/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1194.3042
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/10/2020
Initial Report FDA Received Date01/04/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-