MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) SUSTAIN XL SC; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) SUSTAIN XL SC; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM1134
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Erosion (1750); Unspecified Infection (1930); Pocket Erosion (2013)
Date of Event 12/15/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event or Problem Description
Related manufacturer reference number: 2017865-2021-00144.It was reported that the patient was admitted to the hospital due to pocket rupture and pacemaker exposure.After examination, it was confirmed that the pacemaker pocket was infected and the pocket was ulcerated at the same time.The physician considered that the possible cause of infection is that the patient is too thin.The system was explanted on (b)(6) 2020 and replaced on (b)(6) 2020.The patient was stable.
 
Additional Manufacturer Narrative
Analysis was normal.No anomalies were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUSTAIN XL SC
Common Device Name
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
MDR Report Key11113922
Report Number0002017865-2021-00143
Device Sequence Number3609090
Product Code NVZ
UDI-Device Identifier05414734504744
UDI-Public05414734504744
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P880086
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2020
Device Model NumberPM1134
Device Catalogue NumberPM1134
Device Lot NumberP000078093
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 12/19/2020
Supplement Date Received by Manufacturer01/22/2021
Initial Report FDA Received Date01/04/2021
Supplement Report FDA Received Date02/12/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
TENDRIL ST; TENDRIL ST
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-