• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EMERGE; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EMERGE; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 7133
Medical Device Problem Codes Difficult to Remove (1528); Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/16/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that tip detachment occurred.A 3.00mm x 15mm emerge balloon catheter was prepared, however, the stylet was stuck during removal and the tip broke.The procedure was completed with another of the same device.No complications nor harm were done to the patient.
 
Event or Problem Description
It was reported that tip detachment occurred.A 3.00mm x 15mm emerge balloon catheter was prepared, however, the stylet was stuck during removal and the tip broke.The procedure was completed with another of the same device.No complications nor harm were done to the patient.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: the device was returned for analysis.The shaft, hypotube, tip and balloon were microscopically and visually examined.There was a complete separation at 35.5cm proximal of the tip.There was no proximal portion of the device.The balloon was tightly folded.There were numerous sections of the outer shaft that were stretched down.The inner shaft was stretched for 19mm starting 46mm from the tip.Inspection of the remainder of the device presented no other damage or irregularities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMERGE
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key11114678
Report Number2134265-2020-18487
Device Sequence Number4050182
Product Code LOX
UDI-Device Identifier08714729806233
UDI-Public08714729806233
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K113220
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/22/2023
Device Model Number7133
Device Catalogue Number7133
Device Lot Number0026069901
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/09/2021
Initial Date Received by Manufacturer 12/17/2020
Supplement Date Received by Manufacturer03/10/2021
Initial Report FDA Received Date01/04/2021
Supplement Report FDA Received Date04/02/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-