| Model Number |
7133 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Codes |
No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/16/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that tip detachment occurred.A 3.00mm x 15mm emerge balloon catheter was prepared, however, the stylet was stuck during removal and the tip broke.The procedure was completed with another of the same device.No complications nor harm were done to the patient.
|
| |
|
Event or Problem Description
|
|
It was reported that tip detachment occurred.A 3.00mm x 15mm emerge balloon catheter was prepared, however, the stylet was stuck during removal and the tip broke.The procedure was completed with another of the same device.No complications nor harm were done to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: the device was returned for analysis.The shaft, hypotube, tip and balloon were microscopically and visually examined.There was a complete separation at 35.5cm proximal of the tip.There was no proximal portion of the device.The balloon was tightly folded.There were numerous sections of the outer shaft that were stretched down.The inner shaft was stretched for 19mm starting 46mm from the tip.Inspection of the remainder of the device presented no other damage or irregularities.
|
| |
|