SMITH & NEPHEW, INC. JRNY II BCS ART ISRT TRL SZ 7-8 10MM RT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Model Number |
74035272 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/17/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, during a tka procedure, the both the insert trial and the locking bumper broke during impaction.All pieces recovered, the procedure was successfully completed without delay using a smith+nephew back-up device.No patient injury or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
The device, used in treatment, was returned for evaluation.The clinical/medical team concluded, this case reports that during a tka surgery, both the trial insert and the locking bumper broke during impaction.Per complaint detail, all pieces were recovered from the patient.A photo of the broken devices confirms the pieces were retrieved.Per complaint details, there was no patient injury or delay reported and the procedure was completed using a backup device.Since no patient harm is alleged, no further clinical assessment is warranted at this time.A visual inspection confirmed the jrny ii bcs art isrt trl sz 7-8 10mm rt is broken into pieces.The device shows significant signs of wear/usage.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A review of complaint history for the listed part revealed no prior complaints for the listed batch.A review of the risk management file revealed this failure mode was previously identified.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.This is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.
|
| |
|
Search Alerts/Recalls
|
|
|