MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
|
Back to Search Results |
|
Model Number 97715 |
Device Problems
Computer Software Problem (1112); Overheating of Device (1437); Battery Problem (2885); Charging Problem (2892); Communication or Transmission Problem (2896); Insufficient Information (3190)
|
Patient Problems
Burn(s) (1757); Pain (1994)
|
Event Date 12/31/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Concomitant medical products: product id: 97755, serial#: (b)(4), product type: recharger other relevant device(s) are: product id: 97755, serial/lot#: (b)(4), udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a patient regarding an implantable neurostimulator (ins).It was reported that the ins would not charge, the patient received but doesn't recall specific codes for software problem, the patient also received the screen no device found, and all 0's on passive recharge screen.The controller was also getting hot.The patient stated that they have reset the controller and the situation continued.The patient denied any damage on the recharge telemetry module (rtm) and controller.The rtm coil got very hot and burned their back.The issue was not resolved through troubleshooting.The equipment was not working, and the patient was in "so much pain." the patient mentioned that they were using pain medication to deal with the pain.
|
|
Manufacturer Narrative
|
Continuation of d10: product id: 97755, serial# (b)(6), product type: recharger.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information received reported that the replacement device resolved the reported issues.
|
|
Manufacturer Narrative
|
H3: product id: 97755; serial #: (b)(6); was analyzed and it was found that a "no device found" message was seen, the device was scrapped.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
No new information.
|
|