• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE EXCLUDER AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE EXCLUDER AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number RLT281218
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/10/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The review of the manufacturing paperwork verified that this lot met all pre-release specifications.
 
Event or Problem Description
On (b)(6) 2020, (b)(6) was informed that a patient had a distal type i endoleak, and that a reintervention would take place on (b)(6) 2020.The physician states that the endoleak is likely due to disease progression received additional information from (b)(6).Reintervention took place on (b)(6) 2020.At that time, an excluder extender was implanted, which stopped the endoleak.Patient outcome was good.
 
Additional Manufacturer Narrative
E1: added facility name.G1: revised manufacturing facility.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE EXCLUDER AAA ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
MDR Report Key11117335
Report Number3007284313-2021-01219
Device Sequence Number14433948
Product Code MIH
UDI-Device Identifier00733132622511
UDI-Public00733132622511
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P020004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/25/2021
Device Model NumberRLT281218
Device Catalogue NumberRLT281218
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/11/2020
Supplement Date Received by Manufacturer12/11/2020
Initial Report FDA Received Date01/05/2021
Supplement Report FDA Received Date05/11/2021
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age73 YR
-
-