• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS BIOFLO PORT; PORT & CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS BIOFLO PORT; PORT & CATHETER Back to Search Results
Model Number 44014
Medical Device Problem Codes Fluid/Blood Leak (1250); Fracture (1260)
Health Effect - Clinical Codes Pain (1994); Discomfort (2330); Device Embedded In Tissue or Plaque (3165)
Date of Event 12/26/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The disposable device is not available to be returned to the manufacturer for evaluation.The results of the investigation will be sent via a follow up medwatch.Reference: (b)(4).
 
Event or Problem Description
Approximately 5 months after implantation of a bioflo port, the patient underwent catheter injection under fluoroscopic guidance.On (b)(6) 2019, the patient presented for placement of a bioflo port for the administration of medication of chron's disease.On (b)(6) 2019, the patient underwent a catheter injection under flouroscopic guidance.At this time, the patient experienced severe pain and the injection was terminated.Fluoroscopic findings demonstrated "confined extravasation at the midportion of the catheter, likely into the subcutaneous tissues from catheter defect." on (b)(6) 2020 the patient underwent a procedure to remove the defective device and replace it with another same device.There was no permanent harm or injury to the patient due to this event.The reported device is not available to be returned to the manufacturer for evaluation.
 
Additional Manufacturer Narrative
As the reported defective device was not returned, angiodynamics is unable to perform a device evaluation.The customer's reported complaint description of "catheter tip fractured/migrated" is not confirmed, as no sample was returned.Without receiving product for evaluation, we are unable to definitively determine a root cause for this incident.A review of the device history records was performed for the reported packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.Labeling review: the instructions for use (16608102-01) which is supplied to the user with this catalog number, contains the following statements: contraindications · catheter insertion in the subclavian vein medial to the border of the first rib, an area which is associated with higher rates for pinch-off.Procedure for power injection - absence of a blood return or a poor blood return can be a sign of a potential complication such as occlusion, kinking, breakage, pinch-off syndrome, fibrin formation, thrombosis or malposition.This should be evaluated prior to device usage.A blood return should be present prior to usage of device for any therapy or testing.- if the patient complains of pain, or if there is swelling when the device is flushed or when medication or contrast media is administered, evaluate the device for infiltration, proper needle placement, and potential complications such as occlusion, kinking, breakage, pinch-off syndrome, thrombosis or malposition.Failure to assess these complaints or observations can lead to device failure.Potential complications: catheter fragmentation catheter embolization catheter pinch-off catheter placement considerations: warning: avoid medial catheter placement into subclavian vein through percutaneous technique.This placement could lead to catheter occlusion, damage, rupture, shearing, or fragmentation due to compression of the catheter between the first rib and clavicle.Catheter shearing has been reported when the catheter is inserted via a more medial route in the subclavian vein.Pinch-off syndrome pinch-off syndrome signs may include difficulty in aspirating blood, resistance to flushing or infusion of medications or fluids that improves with position changes, infraclavicular pain and/or swelling with catheter flushing or infusion palpitations, sudden onset chest pain, cardiac arrhythmias, extra heart sound, chest wall swelling at the port pocket, vein insertion site, pain in shoulder or port area not associated with swelling, cough, paresthesia of arm on side of catheter withdrawal occlusion or swishing sound with catheter flushing.Warning: avoid medial catheter placement into subclavian vein through percutaneous technique.This placement could lead to catheter occlusion, damage, rupture, shearing, or fragmentation due to compression of the catheter between the first rib and clavicle.Catheter shearing has been reported when the catheter is inserted via a more medial route in the subclavian vein a review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOFLO PORT
Common Device Name
PORT & CATHETER
Manufacturer (Section D)
ANGIODYNAMICS
10 glens falls technical park
glens falls, ny 12801 NY 12801
MDR Report Key11117504
Report Number1317056-2021-00005
Device Sequence Number1059833
Product Code LJT
UDI-Device IdentifierH965440140
UDI-PublicH965440140
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K131694
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2022
Device Model Number44014
Device Catalogue Number44014
Device Lot Number5474469
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/16/2020
Supplement Date Received by Manufacturer12/16/2020
Initial Report FDA Received Date01/05/2021
Supplement Report FDA Received Date01/21/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-