• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP AG AS VEGA PS FEMORAL COMP.CEMENTED F5N R; KNEE ENDOPROSTHESES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP AG AS VEGA PS FEMORAL COMP.CEMENTED F5N R; KNEE ENDOPROSTHESES Back to Search Results
Model Number NX031Z
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Pain (1994); Joint Laxity (4526)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Reference code: nx031z, device name: as vega ps femoral comp.Cemented f5n r, serial number: n/a, batch number: 51948275, udi device identifier: (b)(4), udi production identifier: (b)(4), basic udi-di: n/a, unit of use udi-di: (b)(4), manufacturing date: 2013-07-25.Reference code: nx053z, device name: as vega ps tibial plateau cemented t2 serial number: n/a, batch number: 51921481, udi device identifier: (b)(4), udi production identifier: (b)(4), basic udi-di: n/a, unit of use udi-di: (b)(4), manufacturing date 2013-02-21.Ref.Code device name batch, nx043 patella 3-pegs p3 51955964, nx121 vega ps gliding surface t2/2+ 12mm 51921827, nn260p plug f/tibial plateau 51967222.Investigation: no product at hand.Therefore an investigation at the product is not possible.Pictorial documentation: there are no pictures available.Batch history review: the device quality and manufacturing history records (dhr) have been checked for all available lot numbers and the products found to be according to our specification valid at the time of production.There are 2 similar complaints against the same lot number 51948275.(all registered as leading component).There are 3 similar complaints against the same lot number 51955964.(all registered as involved component).Explanation and rationale: in the light of the little information received and due the circumstance that we did not received the complained devices it is not possible to determine a root cause for the mentioned failure.Corrective action: for this topic (vega loosening) a product safety case (psc) was initiated.
 
Event or Problem Description
It was reported that there was an issue with a vega knee components.As a result of having the product implanted the patient has experienced pain, swelling in the right knee and normal daily activities were restricted due to pain.Intraoperative findings during replacement surgery were loosening of the tibial and femoral components, which resulted in a revision surgery.The primary surgery occurred on (b)(6) 2013 and the revision surgery occurred on (b)(6) 2018.All available information has been provided at this time, if additional information becomes available the complaint will be updated accordingly.The adverse event / malfunction is filed under reference (b)(4).Involved components: nx031z(ps femur cemented f5n rt), nx043 (universal patella p3), nx053z (ps tibia cemented t2), nx121(ps pe insert t2/t2+, 12mm), nn260p(peek plug f/tibia).Cement used: zimmer palacos r radiopaque bone cement.Associated medwatches: 2916714-2020-00639, 2916714-2020-00640, 2916714-2020-00641, 2916714-2020-00642, 2916714-2020-00643.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AS VEGA PS FEMORAL COMP.CEMENTED F5N R
Common Device Name
KNEE ENDOPROSTHESES
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
MDR Report Key11117932
Report Number2916714-2020-00639
Device Sequence Number19846251
Product Code JWH
UDI-Device Identifier04046963810160
UDI-Public4046963810160
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 01/05/2021
Report Date (Section F) 01/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/01/2023
Device Model NumberNX031Z
Device Catalogue NumberNX031Z
Device Lot Number51948275
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date05/16/2019
Event Location Hospital
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date01/05/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Date Report Sent to FDA01/05/2021
-
-