| Model Number |
406868 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
|
| |
|
Event or Problem Description
|
|
At the end of the procedure, a tear was noted in the body of the sheath when removed from the patient.There were no consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
One 8.5f fast-cath introducer sheath was received for evaluation.The sheath had been torn and stretched; the torn sheath material remained attached.No anomalies were noted when a dilator from current inventory was inserted through the sheath.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported damage remains unknown.The cause of the sheath tear is consistent with damage during use.
|
| |
|