| Model Number |
9-ASD-MF-035 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/09/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The results, method, and conclusion codes along with investigation results will be provided in the final report.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2020, a 35 mm amplatzer cribriform occluder was selected for implant.While deploying, the device wouldn't return to its normal shape.The occluder was recaptured and removed to resolve the event.There was no clinically significant delay in the procedure and the patient remained hemodynamically stable throughout.The patient is stable with no consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event of the device not returning to its normal shape during deployment was confirmed.Following the simulated deployment, the bulbous shape returned to normal and met functional specifications when analyzed at abbott.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all defined manufacturing specifications.A capa was initiated for further investigation and did not identify a product quality issue.However, corrective actions to further enhance performance are being pursued per internal operating procedures.
|
| |
|