| Model Number |
45-20001 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/09/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Upon completion of investigation, additional information will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
As reported: "blade broke while tightening screw.".
|
| |
|
Event or Problem Description
|
|
As reported: "blade broke while tightening screw.".
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could be confirmed, since the device was returned for evaluation and matches the alleged failure mode.The device inspection revealed the following: the received screwdriver shows signs of normal and prolonged usage however its tip is found to be broken.The breakage surface shows relatively smooth surface without any great abrupt features; however, the breakage is slightly uneven.Absence of plastic deformation indicates that the breakage was instantaneous (brittle fracture) due to high torsional and bending overload.The shaft diameter of the screwdriver was observed to be 2.96mm as required against a range of 2.95mm ¿ 3.00mm.Similarly, the average hardness of the screwdriver was observed to be 56.1 hrc as required against a range of 54.0 hrc ¿ 58.0 hrc.As per the dimensional inspection and hardness testing, the returned screwdriver was found to be within specifications.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Based on investigation, the root cause was attributed to be user related.The breakage was caused due to a combination of torsional and bending overload applied on the screwdriver during the surgery.If any additional information is provided, the investigation will be reassessed.
|
| |
|
Search Alerts/Recalls
|