• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD MEDICAL - DIABETES CARE BD MICRO-FINE IV INSULIN SYRINGE; PISTON SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD MEDICAL - DIABETES CARE BD MICRO-FINE IV INSULIN SYRINGE; PISTON SYRINGE Back to Search Results
Model Number 329424
Medical Device Problem Code Packaging Problem (3007)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/09/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The date received by manufacturer has been used for this field.There were multiple lot numbers reported to be involved.The information for each lot number is as follows: medical device lot #: 9119537.Medical device expiration date: 2024-07-31.Device manufacture date: 2019-04-29.Medical device lot #: 9140521.Medical device expiration date: 2024-08-31.Device manufacture date: 2019-05-20.Medical device lot #: 9168926.Medical device expiration date: 2024-09-30.Device manufacture date: 2019-06-17.Medical device lot #: 9189788.Medical device expiration date: 2024-09-30.Device manufacture date: 2019-07-08.Medical device lot #: 9196270.Medical device expiration date: 2024-10-31.Device manufacture date: 2019-07-15.Medical device lot #: 9231122.Medical device expiration date: 2024-11-30.Device manufacture date: 2019-08-19.Medical device lot #: 9287332.Medical device expiration date: 2024-12-31.Device manufacture date: 2019-10-14.Medical device lot #: 9294983.Medical device expiration date: 2024-12-31.Device manufacture date: 2019-10-21.Medical device lot #: 0099954.Medical device expiration date: 2025-06-30.Device manufacture date: 2020-04-08.Medical device lot #: 0160417.Medical device expiration date: 2025-08-31.Device manufacture date: 2020-06-08.Medical device lot #: 8302667.Medical device expiration date: 2024-01-31.Device manufacture date: 2018-10-29.Medical device lot #: 8351757.Medical device expiration date: 2024-03-31.Device manufacture date: 2018-12-17.Medical device lot #: 9322637.Medical device expiration date: 2025-02-28.Device manufacture date: 2019-11-18.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.(b)(4).
 
Event or Problem Description
It was reported that several bd micro-fine¿ iv insulin syringes had issues with partial unsealed packaging units.There were 3 from lot 9119537, 14 from 9140521, 18 from 9168926, 30 from 9189788, 25 from 9196270, 5 from 9231122, 1 from 9287332, 38 from 9294983, 15 from 0099954, 256 from 0160417, 23 from 8302667, 36 from 8351757, and 3 from 9322637.The following information was provided by the initial reporter: "item 329424, a total of 467 units across 13 lot numbers were found to be partly unsealed".
 
Event or Problem Description
It was reported that several bd micro-fine¿ iv insulin syringes had issues with partial unsealed packaging units.There were 3 from lot 9119537, 14 from 9140521, 18 from 9168926, 30 from 9189788, 25 from 9196270, 5 from 9231122, 1 from 9287332, 38 from 9294983, 15 from 0099954, 256 from 0160417, 23 from 8302667, 36 from 8351757, and 3 from 9322637.The following information was provided by the initial reporter: "item 329424, a total of 467 units across 13 lot numbers were found to be partly unsealed".
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d10: device available for eval yes.D10: returned to manufacturer on: 2021-02-22.Investigation summary: customer returned (3) 1cc, 12.7mm, 28g syringes in blister packs from lot # 9119537, 9322637, 0160417, 0099954, 9294983, 9231122, 9196270, 9189788, 9168926, 9140521, 8351757, and 8302667.Customer states that the packaging was not sealed on one side.All returned syringes were examined and all blister packs were observed to be open.A review of the device history record was completed for batch# 9119537, 9322637, 0160417, 0099954, 9294983, 9287332, 9231122, 9196270, 9189788, 9168926, 9140521, 8351757, and 8302667.All inspections were performed per the applicable operations qc specifications.There were zero (0) notifications noted that pertained to the complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MICRO-FINE IV INSULIN SYRINGE
Common Device Name
PISTON SYRINGE
Manufacturer (Section D)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
MDR Report Key11121460
Report Number1920898-2020-01798
Device Sequence Number17128759
Product Code FMF
UDI-Device Identifier30382903294245
UDI-Public30382903294245
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K170386
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number329424
Device Catalogue Number329424
Device Lot NumberSEE SECTION H.10.
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/22/2021
Initial Date Received by Manufacturer 12/09/2020
Supplement Date Received by Manufacturer03/05/2021
Initial Report FDA Received Date01/05/2021
Supplement Report FDA Received Date03/17/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-