• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ARTICULAR SURFACE PROVISIONAL SZ 6 8MM; TEMPLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ARTICULAR SURFACE PROVISIONAL SZ 6 8MM; TEMPLATE Back to Search Results
Model Number 00584102608
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/14/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference case (b)(4).
 
Event or Problem Description
It was reported that, the trial of an articular surface provisional sz 6 8mm is broken and will not seat into tibia trial.Incident occurred during a uka with instruments outside patient.The procedure was successfully completed without delay using a back-up device.No patient injury or other complications were reported.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved or product information, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULAR SURFACE PROVISIONAL SZ 6 8MM
Common Device Name
TEMPLATE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11123190
Report Number1020279-2021-00112
Device Sequence Number666867
Product Code HWT
UDI-Device Identifier00885556619629
UDI-Public00885556619629
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number00584102608
Device Catalogue Number00584102608
Device Lot NumberUNKOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/14/2020
Supplement Date Received by Manufacturer03/16/2021
Initial Report FDA Received Date01/06/2021
Supplement Report FDA Received Date03/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-