| Model Number |
364487 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/26/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Combination product: yes.
|
| |
|
Event or Problem Description
|
|
An orsiro drug-eluting stent system was selected for treatment of a mildly calcified lesion in a moderately tortuous rca.The stent dislodged during crossing of the lesion.The dislodged stent was adapted to the vessel wall with another stent.
|
| |
|
Additional Manufacturer Narrative
|
|
Combination product: yes.The returned product was subjected to a technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for this event.In addition, the angiographic material provided was reviewed.The technical investigation of the returned instrument revealed that the balloon is well folded and shows no signs of inflation.Microscopic analysis revealed stent imprints on the exposed balloon surface, indicating that the stent was properly crimped in between the two radiopaque markers at the time of delivery.The stent was not returned for analysis.The angiographic material shows three stents being successfully implanted from proximal to distal rca.In one of the sequences a dislodged stent is adapted to the vessel wall.The actual complaint event is not visible.Review of the angiographic material did therefore not lead to any further information regarding the nature of the complaint.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in process and final inspections.As a part of final inspection every stent system undergoes visual inspection to ensure correct embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent and the stent retention force of a defined amount of samples is tested from every lot.Based on the conducted investigations, no manufacturing or material related root cause could be determined.
|
| |
|
Search Alerts/Recalls
|