| Model Number |
367004 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/11/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Neither the complaint instrument nor the angiographic material was provided for analysis.Therefore, no technical investigation on the subject could be performed.The production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.In addition to visual inspections each instrument is tested for air tightness by means of a helium leak test.The instrument was delivered in a leak-proof condition.Based on the conducted investigations no manufacturing related root cause could be determined.
|
| |
|
Event or Problem Description
|
|
The pantera leo balloon catheter was used for the post-dilatation of another stent.At first the pantera leo was inflated at the stent middle with 15 atm.During the second inflation at the proximal side of the stent the pantera leo balloon ruptured at 15 atm.
|
| |
|