| Model Number |
391240 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/08/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Combination product: yes.
|
| |
|
Event or Problem Description
|
|
An orsiro drug-eluting stent system was selected for treatment.During stent placement it was noticed that the stent was not on the balloon.The stent was still inside the packaging.
|
| |
|
Additional Manufacturer Narrative
|
|
Combination product: yes.Only the stent was returned and subjected to a technical analysis.The production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for this event.The technical investigation revealed that the stent is severely deformed over its entire length.The delivery system was not returned for analysis.Review of the production documentation for the product detailed above verified that the instrument was manufactured according to specifications and fulfilled all the requirements of in-process and final inspection.During final inspection, every stent system undergoes visual inspection to ensure correct stent embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent and the stent retention force of a defined amount of samples is tested from every lot.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause was determined.It should be noted that, according to the ifu, the stent system should be visually examined to verify functionality prior to procedure.
|
| |
|
Search Alerts/Recalls
|