MAUDE Adverse Event Report: DEPUY SPINE INC VERSE QUICK STICK, TUBE; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SPINE INC VERSE QUICK STICK, TUBE; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 299704217
Medical Device Problem Codes Device-Device Incompatibility (2919); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that during the cleaning process, the quick stick sleeve did not fit properly on the top notch of the screw head.The tip of the inserter appeared stripped.There was no surgical impact.There was no patient consequence.This report is for one (1) verse quick stick, sleeve.This is report 9 of 10 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: h3, h4, h6: a review of the receiving inspection (ri) for verse quick stick, tube was conducted identifying that lot number gm5124504 was released in a single batch.Batch1: lot qty of (b)(4) were released on 21 sep 2018 with no discrepancies.As a result, the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.The ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.Visual inspection: there are a total of 24 verse quick stick, sleeves, and 20 verse quick stick, tubes were received at cq.Upon visual inspection, there were locking facilitator springs noticed bent/deformed on the received devices.Functional testing: all the sleeve and tube combinations were tried to be assembled at customer quality (cq).They were able to be assembled as intended.Complete functional testing could not be performed since the other mating devices(screws) were not returned.However, the deformed/bent condition of the locking spring might have contributed to the reported condition.Can the complaint be replicated with the returned device(s)? unable to perform.Dimension inspection: it was not performed due to post-manufacturing damage.Complaint confirmed? yes.Conclusion: the complaint can be confirmed for unable to assemble as the bent condition could have caused the reported issue.But there was no stripped condition noticed on the device so the complaint would not be confirmed for the stripped issue.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/specification review.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSE QUICK STICK, TUBE
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
MDR Report Key11126896
Report Number1526439-2021-00099
Device Sequence Number3701197
Product Code LXH
UDI-Device Identifier10705034477190
UDI-Public(01)10705034477190
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number299704217
Device Catalogue Number299704217
Device Lot NumberGM5124504
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/29/2021
Initial Date Received by Manufacturer 12/08/2020
Supplement Date Received by Manufacturer01/29/2021
Initial Report FDA Received Date01/06/2021
Supplement Report FDA Received Date02/08/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
VERSE QUICK STICK, SLEEVE; VERSE QUICK STICK, SLEEVE; VERSE QUICK STICK, SLEEVE; VERSE QUICK STICK, SLEEVE; VERSE QUICK STICK, SLEEVE; VERSE QUICK STICK, SLEEVE; VERSE QUICK STICK, SLEEVE; VERSE QUICK STICK, SLEEVE; VERSE QUICK STICK, TUBE
-
-