Model Number S702
Medical Device Problem Code Premature Discharge of Battery (1057)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 11/03/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.A high current drain was detected, but the battery still supported full device function.Detailed analysis confirmed the identified high current drain was a result of compromised capacitors.This resulted in the observed premature battery depletion.It was determined that the capacitors were compromised due to the presence of excess hydrogen in the device case.Boston scientific has issued a field safety notice regarding a subset of pacemakers in the accolade product family that has an elevated potential of exhibiting this behavior.This particular device was not included in the hydrogen induced premature depletion advisory population.However, this capacitor behavior can still occur in non-advisory devices.
 
Event or Problem Description
It was reported that this pacemaker was implanted for less than five years.Premature battery depletion was suspected and the device was explanted and replaced.No additional adverse patient effects were reported.The explanted device was expected to be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALTRUA 2 DR
Common Device Name
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
MDR Report Key11128128
Report Number2124215-2020-27687
Device Sequence Number608017
Product Code LWP
UDI-Device Identifier00802526559341
UDI-Public00802526559341
Combination Product (Y/N)N
Initial Reporter CountryTH
PMA/510(K) Number
N970003/S167
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/11/2017
Device Model NumberS702
Device Catalogue NumberS702
Device Lot Number700059
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/02/2020
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/05/2020
Initial Report FDA Received Date01/06/2021
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-