• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. PDSII VIO 8X18IN 0 S/A CTB-1 CR; SUTURE, SURGICAL, ABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. PDSII VIO 8X18IN 0 S/A CTB-1 CR; SUTURE, SURGICAL, ABSORBABLE Back to Search Results
Model Number ZB740
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/10/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).  a manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.    to date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested and the following was obtained: how the procedure was complete? please provide lot number.Status of product return.After opening the package, the needle tip was broken immediately when holding the needle.When the sales rep received the affected needle, he noticed that there was a mark like blood on the needle.Thus, he asked the hospital if it broke before use on the patient.Then, the hospital assured that it broke before use on the patient.[lot number] unknown: qkmamp.  if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that a patient underwent an unknown surgery on (b)(6) 2020 and the suture was used.During surgery, after opening the package, the needle was broken immediately when holding the needle.The suture broke before use on the patient.There were no adverse patient consequences reported.
 
Additional Manufacturer Narrative
(b)(4).Date sent to the fda: 2/12/2021.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: d9, h6.Additional h3 investigation summary: a suture needle was returned for evaluation.A fracture was observed at the tip of the needle.The needle was received in two pieces, only one of the mating fracture surfaces was examined for this evaluation.A scanning electron microscope (sem) was used to examine the fracture surfaces and surrounding area of the needle.The fracture surfaces were examined in multiple locations to determine the fracture mode.The evaluation of (b)(4) revealed the fracture was composed of microvoid coalescence, which is evidence of a ductile fracture mode.Mechanical damage and a cup-and-cone shaped fracture provide additional evidence that the failure was induced by mechanical deformation leading to ductile overload.This was a ductile fracture.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.The evidence of this examination indicates that the breakage occurred at the tip of the needle during use due to tensile overload.There is no evidence of any material flaw or defect that would cause premature failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PDSII VIO 8X18IN 0 S/A CTB-1 CR
Common Device Name
SUTURE, SURGICAL, ABSORBABLE
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876
MDR Report Key11128214
Report Number2210968-2021-00104
Device Sequence Number846364
Product Code NEW
UDI-Device Identifier10705031062108
UDI-Public10705031062108
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
N18331
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2025
Device Model NumberZB740
Device Catalogue NumberZB740
Device Lot NumberQKMAMP
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/21/2021
Initial Date Received by Manufacturer 12/10/2020
Supplement Date Received by Manufacturer01/21/2021
Initial Report FDA Received Date01/06/2021
Supplement Report FDA Received Date02/12/2021
Patient Sequence Number1
-
-