• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR G4 STEERABLE GUIDING CATHETER; CATHETER, STEERABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR G4 STEERABLE GUIDING CATHETER; CATHETER, STEERABLE Back to Search Results
Catalog Number SGC0705
Device Problems Break (1069); Difficult to Insert (1316); Deformation Due to Compressive Stress (2889)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/18/2020
Event Type  malfunction  
Manufacturer Narrative
The customer reported the device is not returning.Investigation is not yet complete.A follow up report will be submitted with all additional relevant information.
 
Event Description
This is filed to report the steerable guiding catheter (sgc) shaft break.It was reported that this was a mitraclip procedure to treat degenerative mitral regurgitation (mr) with an mr grade of 4.The steerable guiding catheter (sgc), could not be inserted into the inguinal region and the distal shaft broke, kinked and bent after several attempts.The inguinal region was dilated with a balloon and a new sgc was used without further issue.One clip was implanted, reducing mr to 1.There were no adverse patient effects and no clinically significant delay during the procedure.No additional information was provided.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot.Additionally, a review of the complaint history identified no similar complaints reported from this lot.The investigation determined the reported difficulty inserting the device appears to be related to morphology/pathology.The reported broken and kinked shaft were results of the difficulty inserting.There is no indication of a product issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G4 STEERABLE GUIDING CATHETER
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key11128414
MDR Text Key225565914
Report Number2024168-2021-00179
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
PMA/PMN Number
K190167
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/06/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/14/2021
Device Catalogue NumberSGC0705
Device Lot Number01013U114
Was Device Available for Evaluation? No
Date Manufacturer Received02/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age88 YR
Patient Weight62
-
-