• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. BIORAPTOR KNOTLESS SUTURE ANCHOR HIP; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. BIORAPTOR KNOTLESS SUTURE ANCHOR HIP; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Model Number 72202397
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/14/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that during a hip arthroscopy surgery +labral repair, the bioraptor knotless anchor broke off while it was being inserted into the drilled tunnel.There was a surgical delay between 31-60 minutes.Smith and nephew back-up device was available to complete the surgery.All the broken pieces were removed using a grasper.No other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
H3, h6: the reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device history records showed there were no indications to suggest that the lot did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.A review of the instructions for use found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of the raw materials found that a material certificate of analysis is required.A review of risk management files found that the reported failure was documented appropriately.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIORAPTOR KNOTLESS SUTURE ANCHOR HIP
Common Device Name
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
MDR Report Key11128841
Report Number1219602-2021-00068
Device Sequence Number19666838
Product Code MBI
UDI-Device Identifier03596010651280
UDI-Public03596010651280
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K093428
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/10/2025
Device Model Number72202397
Device Catalogue Number72202397
Device Lot Number50855578
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/15/2020
Supplement Date Received by Manufacturer05/03/2021
Initial Report FDA Received Date01/06/2021
Supplement Report FDA Received Date05/04/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-