| Model Number |
PXSLIM045-B |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/09/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has been returned and the investigation results are pending.A follow up mdr will be submitted upon completion of the device investigation.This report is associated with mfr report number: 3005168196-2021-00029.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a coil embolization procedure in the anterior communicating artery (acom) using penumbra coils 400 (pc400s) and a px slim delivery microcatheter (px slim).During the procedure, while advancing a pc400 out of the px slim, the physician experienced resistance and the pc400 would not advance further.The physician then attempted to pull the pc400 in the px slim and experienced resistance again.Subsequently, the pc400 unintentionally detached inside the px slim.Therefore, the detached pc400 and px slim were removed together.The procedure was completed using other coils and another microcatheter.There was no report of an adverse effect to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation of the returned pc400 confirmed that the embolization coil was detached from the pusher assembly and was within the px slim.Further evaluation revealed that the proximal constraint sphere was within the pusher assembly ddt, the sr wire was fractured, and the embolization coil had offset coil winds.If the pc400 is forcefully retracted against resistance, damage such as this may occur.Further evaluation revealed pusher assembly kinks.This damage may be incidental to the reported complaint.Evaluation of the returned px slim revealed an ovalization on the mid-shaft.If the device is forcefully pinched or gripped during use, damage such as an ovalization may occur.This damage likely contributed to the reported resistance experienced during the procedure.Further evaluation revealed a fracture near the hub.This damage was not mentioned in the reported complaint and was likely incidental.Penumbra coils and catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2021-00029.H3 other text : placeholder.
|
| |
|
Search Alerts/Recalls
|