| Model Number |
10371 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Death (1802); Myocardial Infarction (1969)
|
| Date of Event |
06/21/2020
|
|
Type of Reportable Event
|
Death
|
|
Event or Problem Description
|
|
It was reported via social media that death occurred.A left atrial appendage (laa) closure procedure was performed and a 21mm watchman laa closure device was implanted.Nine days post-procedure, the patient had a heart attack and died.
|
| |
|
Event or Problem Description
|
|
It was reported via social media that death occurred.A left atrial appendage (laa) closure procedure was performed and a 21mm watchman laa closure device was implanted.Nine days post-procedure, the patient had a heart attack and died.It was further reported that this was a challenging case as the watchman device did not seat well in the laa.A double curve truseal access sheath was used with a 24mm device initially, but this device would not stay in the laa when tugged.The double curve sheath was then used with a 21mm device hoping to seat it deeper, but it also failed the tug test.The double curve was exchanged for a single curve sheath and a new 21mm device was seated deeper and it passed the release criteria.Both intracardiac echo (ice) and transesophageal echo (tee) were used for this case.
|
| |
|