SMITH & NEPHEW, INC. VULCAN GENERATOR CE MARK; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 7210812
Medical Device Problem Code Failure to Power Up (1476)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/14/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that the vulcan generator was not turning on.This was noticed during set up the procedure; therefore, no patient was involved.A backup device was available and no delay was reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
H3, h6; the reported device, intended for use in treatment, was received for evaluation.A visual inspection was performed on the product and no issue was observed.There was a relationship found between the returned device and the reported incident.Product failed functional testing with power up failure.Cause of power failure is an unplugged main power harness from base board to the power supply.Product passed functional testing after power harness was reseated to power supply connector.The complaint was confirmed and the root cause has been determined to be an unplugged power harness.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VULCAN GENERATOR CE MARK
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key11131390
Report Number1643264-2021-00061
Device Sequence Number14286677
Product Code GEI
UDI-Device Identifier03596010519399
UDI-Public03596010519399
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7210812
Device Catalogue Number7210812
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/28/2020
Initial Date Received by Manufacturer 12/16/2020
Supplement Date Received by Manufacturer01/22/2021
Initial Report FDA Received Date01/07/2021
Supplement Report FDA Received Date01/27/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-