| Model Number |
10601 |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
11/01/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: used the first date of the month of the aware date as no event date was provided.Device evaluated by mfr.: synergy ii us mr 2.25 x 12 mm stent delivery system was returned for analysis.An examination of the crimped stent found stent damage.Stent struts from the mid stent region were noted to be lifted from their crimped position and pulled proximally.The undamaged crimped stent outer diameter was measured and the result was within max crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of damage.A visual and tactile examination of the hypotube shaft found no issues.A visual and tactile examination of the outer and mid-shaft section and a visual examination of the inner lumen found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on (b)(6) 2020.It was reported that stent defect was encountered.A 2.25 x 12 synergy ii drug-eluting stent was selected for use.However, before using the device into the patient, it was noted that stent has a defect on it and could not be used.However, returned device analysis revealed a stent damage.
|
| |
|