U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX LUBRI-SIL ALL-SILICONE FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX LUBRI-SIL ALL-SILICONE FOLEY CATHETER Back to Search Results
Model Number A947316
Medical Device Problem Code Decrease in Pressure (1490)
Health Effect - Clinical Codes Burning Sensation (2146); No Code Available (3191)
Date of Event 12/16/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the catheter was placed and 10cc sterile water was instilled in balloon easily.Later, it was observed that clear yellow urine was draining and patient complaints of burning.The catheter was checked and reporting nurse was unable to remove any saline from bulb.Catheter slid out of patient with bulb completely deflated.
 
Event or Problem Description
It was reported that the catheter was placed and 10cc sterile water was instilled in balloon easily.Later, it was observed that clear yellow urine was draining and patient complaints of burning.The catheter was checked and reporting nurse was unable to remove any saline from bulb.Catheter slid out of patient with bulb completely deflated.
 
Additional Manufacturer Narrative
The reported event was inconclusive because no sample was returned.A potential root cause for this failure could be ¿inflation / drainage lumen wall perforated ¿.It was unknown whether the device had met relevant specifications.The product was used for treatment purposes.It was unknown whether the product had caused the reported failure.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "valve type: use luer slip tip syringe.Do not use needle.Recommended inflation capacities.3cc balloon: use 5cc sterile water.5cc balloon: use 10cc sterile water.30cc balloon: use 35cc sterile water.Do not exceed recommended capacities." h11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text: the device was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX LUBRI-SIL ALL-SILICONE FOLEY CATHETER
Common Device Name
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
MDR Report Key11132153
Report Number1018233-2020-22390
Device Sequence Number846362
Product Code EZL
UDI-Device Identifier00801741074141
UDI-Public(01)00801741074141
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K040504
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/31/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/31/2023
Device Model NumberA947316
Device Catalogue Number175816
Device Lot NumberNGEV1579
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/16/2020
Supplement Date Received by Manufacturer03/31/2021
Initial Report FDA Received Date01/07/2021
Supplement Report FDA Received Date04/23/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-