STRYKER ENDOSCOPY-SAN JOSE PIVOT GUARDIAN, FRAME; APPARATUS, TRACTION, NON-POWERED Back to Search Results
Model Number 3105000100
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/11/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event or Problem Description
It was reported that there was a loss of traction during the procedure.
 
Event or Problem Description
It was reported that there was a loss of traction during the procedure.
 
Additional Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: locking mechanism skipped out and loosened traction.Probable root cause: design: materials bend/dent under force.Design of joints and locking mechanisms.Connectors or locks break under traction force.Casters slip on floor.Boot spine bends.Boot boas unreel due to torque limit.Boa connection guide breaks under high tension.Patient position pad slides on table or patient slides on top of pad.Patient position belt not strong enough to prevent patient from rolling use error.The reported failure mode will be monitored for future reoccurrence.The date of manufacture is not known.H3 other text : 81.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIVOT GUARDIAN, FRAME
Common Device Name
APPARATUS, TRACTION, NON-POWERED
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key11132323
Report Number0002936485-2021-00006
Device Sequence Number2350190
Product Code HST
UDI-Device Identifier07613327380361
UDI-Public07613327380361
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3105000100
Device Catalogue Number3105000100
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/11/2020
Supplement Date Received by Manufacturer12/11/2020
Initial Report FDA Received Date01/07/2021
Supplement Report FDA Received Date03/17/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-