ST. JUDE MEDICAL SWARTZ BRAIDED TRANSSEPTAL GUIDING INTRODUCER, SL1, 63 CM LENGTH, 8.5F; INTRODUCER, CATHETER
|
Back to Search Results |
|
Model Number 407453 |
Device Problems
Flaked (1246); Difficult to Insert (1316)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/16/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported insertion difficulty and skiving could not be conclusively determined.
|
|
Event Description
|
During an atrial fibrillation ablation procedure, while conducting the transseptal puncture, the brk needle could not be introduced completely into the introducer and the inside of the introducer was then noted to be scratched.Moreover, the needle became stuck at the distal end of the introducer each time.The introducer and brk needle were noted to have been prepared correctly and a stylet was used and the needle was not reshaped.The introducer was replaced and the procedure was completed with no adverse consequences to the patient.
|
|