| Model Number |
TMICL12.6 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/01/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The lens was returned in liquied, in a micro-centrifuge vial.Visual inspection found no visible damage to the lens; toric marks are light.Lens work order search-no similar complaint type events reported for units within the same lot.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that the surgeon implanted a 12.6mm tmicl12.6 lens, -13.0/2.0/090 (sphere/cylinder/axis) implantable collamer lens into the patient's eye.The lens was removed and replaced with a backup lens, the surgeon stating he could not see the toric marks on the lens.
|
| |
|
Additional Manufacturer Narrative
|
|
B5-additional information: the lens was implanted, removed, and replaced intraoperatively.H6-health impact code corrected h6-method code 3331, device history record (dhr) review: based on the results of the investigation, all released devices from the associated work order(s), including the suspected device, have been manufactured within established process parameters and there is no indication that the manufacturing and processing of the device contributed to the complaint issue.Claim# (b)(4).
|
| |
|