Model Number MD-631
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/09/2020
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus medical systems corp.(omsc) for evaluation.Omsc will start evaluating the device.There were no further details provided.If significant additional information is received, this report will be supplemented.
 
Event Description
During the hysteroscopic surgery with clv-s190 using the device (xenon lamp md-631), there was a bursting sound and the device went off.The emergency lamp of the clv-s190 stated up.The user completed the procedure by using the clv-s190.There was no report of patient injury associated with the event.The device was broken.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.The subject device was returned to olympus medical systems corp.(omsc) for evaluation.Omsc confirmed cracks over the entire surface of the lens of the subject device.Omsc confirmed that four years or more have passed since the subject device was manufactured.The exact cause of the reported phenomenon could not be conclusively determined.However, it is surmised that the reported phenomenon was attributed to an accidental failure or over heated due to dust on the heat exhaust of the light source.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XENON LAMP
Type of Device
XENON LAMP
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11136527
MDR Text Key238350665
Report Number8010047-2021-01051
Device Sequence Number1
Product Code GCT
Combination Product (y/n)N
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 02/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/07/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMD-631
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/10/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-