The manufacturer previously reported receiving information alleging that a ventilator alarmed for exhalation valve failure while on a patient.The patient developed respiratory insufficiency and required respiratory resuscitation with a manual bag resuscitation device.The patient was transferred to the intensive care unit and required prolonged hospitalization due to the event.The device was returned to the manufacturer's quality product investigation lab for evaluation.During the evaluation, a "service required" code alarm condition related to the device's active exhalation valve was observed in the ventilator's downloaded event log.An examination of the device's active exhalation valve revealed excess fluid in the diaphragm and tubing of the active exhalation valve.The manufacturer was able to confirm the complaint of active exhalation valve failure.The manufacturer concludes the root cause of the active exhalation valve failure is excess water in the device's active exhalation valve diaphragm and tubing.The excess water likely prevented the active exhalation valve from opening and closing properly.The manufacturer also received the circuit and valve that were used during this event.The f&p circuit and intersurgical valve that were returned with the device are not on the manufacturer's list of approved accessories for use with the trilogy evo device.A review of the ventilator's downloaded event log showed multiple patient circuit and error alarms on the date of the reported event which occurred from the time the device was powered on.These alarms were either not acknowledged or were silenced as evidenced in the device's downloaded event logs.The trilogy evo device was evaluated and found to operate and alarm as designed on the manufacturer's multifunction test station.The device passed all testing and performed to specifications.The manufacturer has determined that the trilogy evo device works as designed but when the device is connected to the patient circuit that was returned with the device, the performance of the device is diminished due to the fluid in the diaphragm and tubing which causes the failure.In addition to the above information, a "service required" code was found in the ventilator's downloaded error log related to the device's system board.The device's system board needs replaced to address the issue.
|