| Model Number |
9736060 |
| Medical Device Problem Code |
Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/11/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a guidance system being used outside of a procedure.It was reported that the instrumentation trays were damaged and the instruments failed tracking during preventative maintenance.The trajectory error ranged between 0.554 and 0.976 which was above the maximum amount allowed.Troubleshooting involved trying different spheres, trying different navlock trackers and snapshot tracker, using long and short tools, adjusting the mounted position of the surgical system, hard restarts of the guidance system and trying different surgical arms.The stealth camera was viewing all of the spheres well during testing.The most probable cause was noted to be the washing and cleaning by the sterile processing department.The instruments were in rough condition and the edges.The edges of the trackers were chipped and the paint was faded.The trackers were difficult to detect during testing.There was no patient involvement.
|
| |
|
Event or Problem Description
|
|
Additional information received from a manufacturer representative reported that the issue was determined to be the camera.The camera was replaced and the issue was resolved.All of the instruments were tested and found to be functioning as designed.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|