• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIFEWATCH SERVICES, INC. LIFEWATCH MOBILE CARDIAC TELEMETRY MCT 3-LEAD ( UNIVERSAL GATEWAY)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIFEWATCH SERVICES, INC. LIFEWATCH MOBILE CARDIAC TELEMETRY MCT 3-LEAD ( UNIVERSAL GATEWAY) Back to Search Results
Model Number MCT 3L
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Hypersensitivity/Allergic reaction (1907)
Date of Event 11/26/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Per patient user guide sup546, patients are provided warning statements regarding electrode skin irritation.Patients are instructed to contact lifewatch services customer support prior to using the mct 3l monitoring system if they have known allergies to nickel or other metals.The patient is also informed in the user guide that if they develop skin irritation they should contact their healthcare professional.Specific patient intrinsic, extrinsic, and preventable factors are unknown.No previously known skin sensitivities or allergies were reported by the patient.The patient allegedly consulted a clinician and was prescribed prednisone.The patient was also offered alternative electrodes that are more suitable for sensitive skin.The ifu located on the pouch states the electrodes are hypoallergenic and an electrode is to be worn for less than 72 hours.The electrodes are also pvc and latex free.The electrode supplier has not undergone changes in the production process or raw materials and meet requirements for skin sensitivity / cytotoxicity / irritation.No additional information is known to braemar manufacturing, llc at this time.
 
Event or Problem Description
Patient communication of consulting a healthcare professional due to allergic reaction/skin irritation where treatment was prescribed.
 
Event or Problem Description
Patient communication of consulting a healthcare professional due to allergic reaction/skin irritation where treatment was prescribed.
 
Additional Manufacturer Narrative
Per patient user guide sup546, patients are provided warning statements regarding electrode skin irritation.Patients are instructed to contact lifewatch services customer support prior to using the mct 3l monitoring system if they have known allergies to nickel or other metals.The patient is also informed in the user guide that if they develop skin irritation they should contact their healthcare professional.Specific patient instrinsic, extrinsic, and preventable factors are unknown.No previously known skin sensitivities or allergies were reported by the patient.The patient allegedly consulted a clinician and was prescribed prednisone.The patient was also offered alternative electrodes that are more suitable for sensitive skin.The ifu located on the pouch states the electrodes are hypoallergenic and an electrode is to be worn for less than 72 hours.The electrodes are also pvc and latex free.The electrode supplier has not undergone changes in the production process or raw materials and meet requirements for skin sensitivity/cytotoxicity/irritation.Nissha medical technologies/graphic controls investigated the event, including a review of the lot record paperwork, quality plans, and associated processes.This review found no nonconformances or anomalies.A trend report showed that there have not been any significant increases in skin reaction complaints over the past three years (lifewatch customer complaint reference cas - (b)(4)).The patient also used alternate cardinal health electrodes during the course of the enrollment.The supplier was notified of the event and lifewatch is awaiting supplier response regarding their evaluation.(lifewatch customer complaint reference cas - (b)(4)) no additional information is known to braemar manufacturing, llc at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEWATCH MOBILE CARDIAC TELEMETRY MCT 3-LEAD ( UNIVERSAL GATEWAY)
Common Device Name
LIFEWATCH MOBILE CARDIAC TELEMETRY MCT 3-LEAD ( UNIVERSAL GATEWAY)
Manufacturer (Section D)
LIFEWATCH SERVICES, INC.
10255 w higgins rd.
suite 700
rosemont IL 60018 5606
MDR Report Key11143062
Report Number3027765-2021-00001
Device Sequence Number2994882
Product Code DSI
UDI-Device Identifier00850725007101
UDI-Public00850725007101
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K143359
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberMCT 3L
Device Catalogue NumberMCT 3L: UNIVERSAL GATEWAY
Device Lot Number3971V13
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/10/2020
Supplement Date Received by Manufacturer12/10/2020
Initial Report FDA Received Date01/08/2021
Supplement Report FDA Received Date09/02/2021
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age59 YR
-
-