MAUDE Adverse Event Report: WARSAW ORTHOPEDICS ZEVO ANTERIOR CERVICAL PLATE SYSTEM; APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WARSAW ORTHOPEDICS ZEVO ANTERIOR CERVICAL PLATE SYSTEM; APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY Back to Search Results
Model Number 3001027
Medical Device Problem Codes Device Dislodged or Dislocated (2923); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/22/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from healthcare professional via field representative regarding patient with unknown indication involved in acdf procedure.Levels implanted: c4-5.It was reported post- operatively, the locking mechanism bent and allowed the screw to back out.Implant date is unknown.A revision surgery was completed on (b)(6)2020, to remove the bent plate and old screws.Product was utilized correctly according to the directions given in the ifu/labeling.Patient is alive and no injury.
 
Additional Manufacturer Narrative
H10: radiographic image review result a single post-op ct image is provided procedure listed as acdf c4-5.One of the screws is backed out of the vertebral body and plate.The level appears to be c7 or t1 given the presence of rib heads on the ct.Fusion status is unknown.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Device evaluated summary- visual and optical inspection confirmed one of the locking tabs has been bent and not laying flat against the plate.The tab will no longer stay locked.The tab may have bent when the screw was threaded into the plate.This type of damage is consistent with bend stress overload.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZEVO ANTERIOR CERVICAL PLATE SYSTEM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY
Manufacturer (Section D)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
MDR Report Key11146149
Report Number1030489-2021-00039
Device Sequence Number846057
Product Code KWQ
UDI-Device Identifier00643169329881
UDI-Public00643169329881
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
K141632
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3001027
Device Catalogue Number3001027
Device Lot Number0353602W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/15/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/22/2020
Supplement Date Received by Manufacturer02/02/2021
04/30/2021
Initial Report FDA Received Date01/11/2021
Supplement Report FDA Received Date02/11/2021
05/03/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-