Model Number 3L92514
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Adhesion(s) (1695); Insufficient Information (4580)
Date of Event 12/24/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Hcp is reporting a fracture of a corail stem at taper area.Dor: (b)(6) 2020.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative:  added: d9.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: d3, g1, h3.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : examination of the returned device and attached photographs confirmed the reported fracture of the stem.The root cause is attributed to the hip stem being cyclically loaded beyond the material limit.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot : a review of the dhr has been completed and no deviations or anomalies were found.
 
Event or Problem Description
Patient received a primary right tha in (b)(6) 2016.The cup, head, and liner were competitor products implanted with a depuy corail stem.Primary operative notes were not provided.Revision 1: patient received a right hip revision of a competitor femoral head to treat pain, limb asymmetry (shortening of the right leg), and mobility limitations of the right hip.Upon entering the joint, the surgeon encounters and excises extensive periarticular ossifications and scarring.The surgeon notes that intraoperative luxation of the joint is difficult due to flexion contracture.The competitor head is removed and replaced with a competitor revision ceramic head with and xl length adapter with 7.50 valgus positioning to treat the reported limb asymmetry.The cup and polyethylene liner are well-positioned and without wear and are retained.Local radiation was also performed as a prophylaxis of periarticular ossifications.The procedure was completed without complications.This procedure was previously unreported.Revision 2: patient received a right hip femoral revision to treat fatigue fracture of the femoral head/neck junction.Upon entering the joint, the surgeon notes dense scarring.The stem is well-fixed and requires an extensive osteotomy to loosen and remove the device in order to revise the broken shaft portion with the competitor femoral head/neck prosthesis.Once the stem is removed, the osteotomy fragments are fixed with cerclage and a screw and cemented competitor products are implanted with a depuy cement plug.The procedure was completed without complications.Doi: (b)(6) 2016 dor: (b)(6) 2017- revision 1 of competitor head dor: (b)(6) 2020- revision 2 of competitor head and corail stem right hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORAIL2 STD SIZE 14
Common Device Name
CORAIL AMT CEMENTLESS IMPLANT : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY FRANCE SAS - 3003895575
7 allee irene joliot-curie
b.p. 256
saint priest cedex 69801
FR  69801
MDR Report Key11146906
Report Number1818910-2021-00766
Device Sequence Number630643
Product Code KWA
UDI-Device Identifier10603295168805
UDI-Public10603295168805
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K042992
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 12/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3L92514
Device Catalogue Number3L92514
Device Lot Number5255147
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/19/2021
Initial Date Received by Manufacturer 12/28/2020
Supplement Date Received by Manufacturer01/16/2021
02/19/2021
03/25/2021
Initial Report FDA Received Date01/11/2021
Supplement Report FDA Received Date02/05/2021
02/23/2021
03/25/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN HIP FEMORAL HEAD; UNKNOWN HIP FEMORAL HEAD
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR
-
-