Model Number N/A
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 12/08/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Medical product: catalog #: unknown, reverse tray, lot # unknown; catalog #: 113627, comp primary stem 7mm mini mini, lot # 488500.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted multiple mdr reports were filed for this event, please see associated reports: 0001825034 - 2020 - 04382.
 
Event or Problem Description
It was reported that during a comprehensive reverse total shoulder procedure a humeral stem could not be attached to the inserter.The stem was inserted by hand, and then the humeral tray would not engage the stem.The stem was then removed.Additional broaching was required and a larger size stem was used to complete the procedure.
 
Additional Manufacturer Narrative
(b)(4).Upon receipt of additional information it has been determined that this device did not cause or contribute to the reported event.No serious injury or harm to the patient reported for this event or prior events with same or similar products.The initial report was forwarded in error and should be voided.
 
Event or Problem Description
Upon receipt of additional information it has been determined that this device did not cause or contribute to the reported event.No serious injury or harm to the patient reported for this event or prior events with same or similar products.The initial report was forwarded in error and should be voided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMP PRIMARY STEM INSERTER
Common Device Name
PROSTHESIS SHOULDER
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key11146917
Report Number0001825034-2021-00025
Device Sequence Number3286060
Product Code LXH
UDI-Device Identifier00880304456648
UDI-Public(01)00880304456648(11)180514(10)287480
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number407398
Device Lot Number287480
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 12/22/2020
Supplement Date Received by Manufacturer05/06/2021
Initial Report FDA Received Date01/11/2021
Supplement Report FDA Received Date05/07/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
-
-