• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P BIOLOX DELTA HIP; HEAD, FEMORAL, CERAMIC, BILOX DELTA, 32MM NEUTRAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P BIOLOX DELTA HIP; HEAD, FEMORAL, CERAMIC, BILOX DELTA, 32MM NEUTRAL Back to Search Results
Model Number 400-03-322
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 12/14/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Event or Problem Description
Revision surgery - patient was brought back to the or for dislocation of the hip.Surgeon changed out femoral head and proximal body during the procedure.
 
Additional Manufacturer Narrative
Manufacturer narrative: the reason for this revision surgery was reported as dislocation.The previous surgery and the surgery detailed in this event occurred 1 year and 5 months apart.The healthcare professional indicated that there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at the hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was revision surgery due to dislocation.There were no findings during this evaluation that indicate the reported devices were defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may also contribute to an event that are outside the control of djo surgical such as poor bone density, patient bone deterioration, inadequate soft tissue support, patient activities or trauma.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Event or Problem Description
Second revision surgery - patient was brought back to the operatin room for dislocation of the hip.Surgeon changed out femoral head and proximal body during the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOLOX DELTA HIP
Common Device Name
HEAD, FEMORAL, CERAMIC, BILOX DELTA, 32MM NEUTRAL
Manufacturer (Section D)
ENCORE MEDICAL L.P
9800 metric blvd
austin,tx 78758-5445
MDR Report Key11149898
Report Number1644408-2020-01260
Device Sequence Number16335438
Product Code LZO
UDI-Device Identifier00888912076098
UDI-Public(01)00888912076098
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K082844
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/07/2022
Device Model Number400-03-322
Device Catalogue Number400-03-322
Device Lot Number861B1103
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/14/2020
Supplement Date Received by Manufacturer03/11/2021
Initial Report FDA Received Date01/11/2021
Supplement Report FDA Received Date04/08/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
495-00-065 LOT 539U1062
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
-
-