| Model Number |
SBC-501 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/14/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown.(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a seal biopsy valve was used during a colonoscopy procedure performed on (b)(6) 2020.According to the complainant, during the procedure after passing pediatric forceps down the scope channel the physician noticed a foreign debris inside the patient.A net was used to retrieve the foreign material.They took off the biopsy valve and noted that the debris was from the biopsy valve being fractured.The procedure was completed using the original seal biopsy cap.There were no patient complications reported as a result of this event.
|
| |
|