| Model Number |
7440020 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Device Slipped (1584)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/23/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Mfr site: country: (b)(4).Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Information was received from a healthcare provider via manufacturer representative regarding a patient with an indication of l3 cone fracture in need of anterior l3 cone resection, spinal canal decompression, bone graft pedicle screw fixation and fusion used in spinal therapy.It was reported that the set screw slipped and couldn't be fixed.There were no patient symptoms reported.There were no fragments in the patient reported.There was no delay in procedure time reported.There were no further complications reported regarding the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: visual and optical examination revealed thread crest/flank damage; this damage appears to have initiated at or near the start of the thread and is evident around the damaged portion of the thread.Functional evaluation with sample bone screw revealed the set screw was unable to be fully engaged.The above observations are consistent with misalignment of the bone screw and set screw threads during construct assembly.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|